Lasers Used in Thermal Laser Stimulation
- Practical Electron Microscopy and Database -
- An Online Book -

https://www.globalsino.com/EM/  



 
This book (Practical Electron Microscopy and Database) is a reference for TEM and SEM students, operators, engineers, technicians, managers, and researchers.
 

=================================================================================

The lasers with 1340 nm wavelength eliminate the LIVA (light-induced voltage alteration) signal because the energy of the photons is below that of the Si (silicon) indirect band-gap. The photocurrent effects from e-h (electron-hole) pair generation by the 1064 nm laser used in LIVA produce signals 1 to 2 orders of magnitude larger than the SEI and TIVA signals so that the 1064 nm laser should be avoided in SEI and TIVA measurements. Note that some lasers with 1340 nm wavelength still contain a weak intensity (e.g. < 1%) of light with 1047 nm wavelength; therefore, a long wavelength pass filter needs to be placed in the beam path in order to eliminate the unwanted LIVA signals.

 

 

=================================================================================