EELS Energy Offset - Practical Electron Microscopy and Database - - An Online Book - |
||||||||
Microanalysis | EM Book https://www.globalsino.com/EM/ | ||||||||
This book (Practical Electron Microscopy and Database) is a reference for TEM and SEM students, operators, engineers, technicians, managers, and researchers. | ||||||||
================================================================================= | ||||||||
In EELS acquisition, if there is a nonzero offset, the spectra can be too weak to be visible. Therefore, the Spectrum Offset needs to be turned off or set to 0. If the EELS Energy Offset does not work, we need to check: We need to recalibrate the condenser lens adjustment from the FilterControl software if the EEL specimen illumination changes when the spectrum offset is changed.
|
||||||||
================================================================================= | ||||||||