Conditions to Minimize Electron-Beam-Induced
Reduction of Materials in EELS Measurements
- Practical Electron Microscopy and Database -
- An Online Book -

https://www.globalsino.com/EM/  



 

This book (Practical Electron Microscopy and Database) is a reference for TEM and SEM students, operators, engineers, technicians, managers, and researchers.

 

=================================================================================

For instance, to minimize the electron-beam-induced chemical reduction of Mn4+ in Mn-containing minerals [1], each analysis has to be obtained in a low-dose condition using the lowest probe current while still producing a sufficient signal-to-noise ratio. This damage minimization can be done by spreading the electron beam for a low irradiation rate, e.g. smaller than 2 × 104 e/nm2/s. At this condition, the acquisition time for each spectrum can be relatively long, e.g. 8–10 s. If the counts for a single spectrum are not sufficient to produce pronounced sharp peaks, a series of spectra should be summed to improve counting statistics such that the integrated dose for each accumulated spectrum is more than 2 × 106 e/nm2.

 

[1] Garvie, L.A.J. and Craven, A.J., (1994) Electron-beam-induced reduction of Mn4+ in manganese oxides as revealed by parallel EELS. Ultramicroscopy, 54, 83–92.

 

=================================================================================

The book author (Dr. Liao) welcomes your comments, suggestions, and corrections, please click here for submission. You can click How to Cite This Book to cite this book. If you let Dr. Liao know once you have cited this book, the brief information of your publication will appear on the “Times Cited” page.



 
 
 