Chapter/Index: Introduction | A | B | C | D | E | F | G | H | I | J | K | L | M | N | O | P | Q | R | S | T | U | V | W | X | Y | Z | Appendix
| The intensity of the transmitted X-ray is a function of the intensity of a generated X-ray (I0), mass absorption coefficient μ/ρ (cm2/g), material density ρ (g/cm3), and thickness (cm): As an example, page1730 shows X-ray absorption and intensity reduction induced by carbon contamination in EMs. Equation 1313a can be re-written as, Assuming all X-rays, at different wavelengths, penetrating the same materials through the same thickness, e.g. X-rays collected by EDS detector in TEM or SEM systems, then the collected X-ray intensities decreases experientially depending on the corresponding mass absorption coefficient as indicated by Equation 1313b. Table 1313. Mass absorption coefficients (μ/ρ, cm2/g) of several elements.
The mass absorption coefficient (e.g. Figure 1313a) can be given by,
The corresponding mass absorption coefficient of a compound can be calculated by the equation, Note that the mass absorption coefficient for a mixture can also be calculated in the same way as shown in Equation 1313c.
|