EDS Energy Calibration
- Practical Electron Microscopy and Database -
- An Online Book -
Electron microscopy
  Microanalysis | EM Book                                                                   https://www.globalsino.com/EM/  
 

 

The peak positions of EDS spectrometers should be calibrated so that the positions of characteristic X-ray lines are found within 10 eV of the tabulated values. A miscalibrated system will shift X-ray peaks to improper energies, increasing the risk of misidentification. On the other hand, due to the possible drift of the electronics of the EDS systems, the calibration should be checked periodically to ensure the quality of the EDS measurements.

Calibration can usually be done on pairs of two widely separated X-ray lines, for instance:
         i) Overlapping Al Kα lines at ~1.486 keV (namely, with Kα2 at 1.486 keV and Kα1 at 1.487 keV) and overlapping Cu Kα lines at ~ 8.037 keV (namely, with Kα2 at 8.027 keV and Kα1 at 8.047 keV).
         ii) Overlapping Cu Lα2,1 at 0.928 keV and overlapping Cu Kα lines at ~ 8.037 keV.
         iii) Overlapping Cu Kα lines at ~ 8.037 keV and the zero-strobe peak.

For EDS systems on SEMs, the calibration may require increasing the accelerating voltage of the electron beam, and thus the higher energy X-rays can be generated and recorded.

;