Elements with Possible X-ray Interferences in EDS
- Practical Electron Microscopy and Database -
- An Online Book -

https://www.globalsino.com/EM/  



 
This book (Practical Electron Microscopy and Database) is a reference for TEM and SEM students, operators, engineers, technicians, managers, and researchers.
 

=================================================================================

Some X-ray lines are so close that special care may be required. Tables 2516 lists the elements with possible X-ray interferences in EDS.

Table 2516. Elements with possible X-ray interferences in EDS.

Elements and lines Energy (eV) Elements and lines Energy Elements and lines Energy
V Kα 4,952.2 Mo Lα 2,293.2 Cr Kα 5,414.7
Ti Kβ 4,931.8 S Kα 2,307.8 V Kβ 5,427.3
F Kα 676.8 Pb Lα 10,551.5 Mn Kα 5,898.8
Fe Lα 705.0 As Kα 10,543.7 Cr Kβ 5,946.7
Si Kβ 1,835.9 Re Lα 8,652.5 Fe Kα 6,403.8
Sr Lα 1,806.6 Zn Kα 8,638.9 Mn Kβ 6,490.4
Zn Lα 1,011.7 Ba Lβ 4,466.3 Ga Kβ 10,264.2
Na Kα 1,041.0 Ce Lα 4,840.2 As Kα 10,543.7
Zn Lα 1,011.7 N Kα 392.4 Zr Kβ 17,667.8
Cu Lα 929.7 Ti L(l) 395.3 Mo Kα 17,479.3

 

 

 

 

=================================================================================

The book author (Yougui Liao) welcomes your comments, suggestions, and corrections, please click here for submission. If you let book author know once you have cited this book, the brief information of your publication will appear on the “Times Cited” page.