Practical Electron Microscopy and Database

An Online Book, Second Edition by Dr. Yougui Liao (2006)

Practical Electron Microscopy and Database - An Online Book

Chapter/Index: Introduction | A | B | C | D | E | F | G | H | I | J | K | L | M | N | O | P | Q | R | S | T | U | V | W | X | Y | Z | Appendix

FEI TEM Operation Manuals

 
TEM Operation Procedure:
TEM models
  • Fill the Dewar flask with liquid nitrogen (LN2). If the Cold Trap is at room temperature, allow it to cool for 20~30 minutes, which may be taken before the column vacuum reaches the necessary level. Replenish liquid nitrogen as necessary during the operation session.
  • Launch the TF20 TEM User Interface (UI):
    • Open the User Interface (UI) through the Windows quick launch toolbar or the Start menu.
    • Navigate to Workset → Setup → FEG Control (User). Verify that the FEG Control parameters are configured as follows: HT = 200kV, Extraction Voltage = 3800V, and Gun Lens (GL) = 4. If you need to adjust the extraction voltage, input the desired value and then press the Enter button located next to the field.
    • Navigate to Workset → Setup → Vacuumto confirm the vacuum levels in the system:
      • The gun vacuum should always read 1.
      • The column vacuum should be below 20 log or <10-7 Torr).
      • The Col. Valves Closed button is illuminated yellow, indicating that the valves are closed. The column valves must remain closed if the column vacuum exceeds 20 log or when the microscope is idle. Opening the column valves under high vacuum conditions can severely damage the field emission gun.
    • Launch the digital camera control software, e.g. TIA.
  • (Optional) Load the latest alignment by navigating to Workset > Alignment > File.
  • Set the most recent FEG Registry from FEG Registers. In Microscope control, select the tab; navigate down to the “FEG Registers” control panel. [11]
  • Set the most recent FEG Registry from FEG Registers

  • Choose a sample holder. [8]

    Choose a sample holder

  • Follow the correct procedure to insert the specimen holder into the microscope column.
  • If the column log is 20 or lower, open the gun valve and the column valves by clicking the Col. Valves Closed button (see Vacuum). The button will change from yellow to gray, and the Setup tab will display the status: Column Opened. When the column valves are opened, the CompuStage light will glow red. Avoid touching the CompuStage or specimen holder while the red light is illuminated!

    Valve status

    Figure 0364h. Valve status. The button will turn gray, the turbo pump will turn off, and status will read “READY”. [11]
  • Adjust the microscope settings to the most commonly used configurations:
    • Magnification: 2,000x - 5,600x.
    • Spot size: 3–4.
    • C1 aperture: 2000 mm.
    • C2 aperture: 100 mm.
  • Reset the holder by navigating to Workset → Stage → Stage2 → Control → Reset → Holder. Resetting the holder at the beginning and end of a microscope session is recommended. On the other hand, if issues arise with controlling the CompuStage, resetting the holder can sometimes resolve the problem.
  • Position and align the objective and/or selected area aperture(s) at the center if needed.
  • Press the “Eucentric focus” button and normalize all (L1 or L2) button (see L3R3).
  • Adjust the Z-height: Identify a distinctive feature and center it on the screen. Navigate to Stage → Stage2 → flap-out, and then Control-click the Wobbler button. When activated, the button will turn yellow, causing the goniometer to oscillate continuously between the preset + and – angles. This movement will create a back-and-forth image shift. Reduce the amplitude of this motion by using the Z-axis buttons until the movement is minimized. Finally, turn off the Wobbler and verify that A&B values are close to “0.”
  • Repeat "Adjust the Z-height" at higher magnification (~50,000x) to fine tune the Z position.
  • For the step of "Adjust the Z-height", if your sample is crystalline, instead of using Wobbler function menioned above, you can focus the beam and adjust the Z-height up or down until the diffraction pattern aligns with the central spot.
  • View the image on the screen and adjust the intensity to evenly spread the beam across it.
  • Alignments.
 
The interface of a microscope contains three parts:
  • TEM Server. It controls the entire microscope. Since the TEM Server plays a critical role in the operation of the TF20, it is restricted from access by general users.
  • User Interface.
    • Users interact with the TEM Server—and consequently the microscope—through the User Interface (UI), which serves as a secondary application layer. The UI enables access to essential microscope functions, with control panel knobs and buttons designed to simulate traditional TEM controls. In practice, these controls interface with the computer, which manages all microscope operations.
    • The UI includes various panels and windows. The central area is allocated for supplementary programs such as TIA and others. The lower section of the user interface primarily displays information panels, such as the Message panel, Status panel showing current microscope parameters, and others. [1] The vertical section of the UI includes multiple control panels for managing various microscope parameters. Each control panel includes a collection of controls associated with a primary microscope function, such as Field Emission Gun (FEG) control, CompuStage (stage control), camera control, and more. All of these control panels are grouped into a Workset and can be accessed by selecting the corresponding tab at the top. Drop-downs are used to expand the list of available options or choices on certain control panels. The Workset serves as an organizer for commonly used microscope controls and facilitates access to the Control panels. A complete list of available microscope controls can be accessed through the Popup panel located at the bottom-right corner of the UI. The Popup panel allows users to open Control panels (alongside the Workset) and manage items in the Workset, including adding, removing, or renaming them.

      User Interface

      Figure 0364a. Microscope User Interface. Adapted from [1]

      F1 function key: You can press the F1 function key to access the UI help pages from any Control panel. Simply position the cursor over a specific item and press F1, and the corresponding help pages for that item or Control panel will appear automatically.

      • The Vacuum Control panel displays the vacuum levels, e.g. listed in Table 0364, in various sections of the microscope, such as the gun and the column. [2] The column valves should remain closed when the microscope is not in use or if the vacuum reading in the column exceeds 20 Log or 10-7 Torr. A yellow button indicates that the valves are currently closed, while a gray button signifies that the function or device is not active. When the "Column Valves Closed" button appears yellow, the valves are securely closed. Clicking this button changes its color to gray, indicating that the column valves have been opened.

        Table 0364. Typical values (log) for the FEI TEM vacuum system.[1-12]

        Gun or Accelerator IGP3 1
        Column IGP1 1-6
        Detection limit   13
        Buffer tank   53
        Camera P3 20
        Backing line   88
        Liquid nitrogen  If applicable > 10%

        The Vacuum Control panel displays the vacuum levels in various sections of the microscope, such as the gun and the column The Vacuum Control panel displays the vacuum levels in various sections of the microscope, such as the gun and the column

        Figure 0364b. Vacuum Status. [1, 6]

        This control panel below can be accessed exclusively through the Popup panel located in the bottom-right corner of the user interface. Various colors indicate the quality of the vacuum: cyan signifies an optimal vacuum level for the specific component associated with that color.

        • GP – Ion-Getter Pump.
        • TMP – Turbo-Molecular Pump.
        • Pir, Pen – vacuum gauges.
        • V - valves
        • .

          The Vacuum Control panel displays the vacuum levels in various sections of the microscope, such as the gun and the column

          The Vacuum Control panel displays the vacuum levels in various sections of the microscope, such as the gun and the column

      • At 200kV, the extraction voltage is set to 3800V, and the gun lens is adjusted to 4 (within the range of 3 to 6). A yellow Operate button signifies that the FEG is active, while a yellow High Tension button indicates that the high voltage is enabled.

        The Vacuum Control panel displays the vacuum levels in various sections of the microscope, such as the gun and the column

    • EDS acquisition and analysis (Esprit: Bruker spectrometer system, and Velox)
    • .
    • The level of control varies based on the user's position within FEI's user-account hierarchy: factory → service → supervisor → user. While users have sufficient control to operate the microscope across all available modalities, their ability to alter critical TF20 settings or adjustments is restricted.
    • Each user has their own standard Windows account secured with a password.
    • Each new account adopts the supervisor settings active at the time of its creation. These settings can later be modified and saved by the user within their account.
    • Users can access and apply settings from others, including those of the supervisor, highlighting the benefits of a digital interface.
  • Control-panels (left and right) shown in the figure below [1]. The control panels include a typical arrangement of TEM controls:
    • Intensity (brightness).
    • Magnification.
    • Focus.
    • Focus-step.
    • Dark field.
    • Diffraction.
    • Wobbler.
    • Stage X-Y joystick.
    • Beam shift trackball.
    • Z-height.
    • Programmable buttons. The system includes a set of user-defined programmable buttons that operate as switches. These buttons can be customized to perform frequently used functions, such as adjusting the screen lift or spotsize.
    • Multifunction knobs X and Y (MF).
      • The default function of the MF knobs is "user beam shift X & Y."
      • They can also be linked to other functions, such as beam tilt, astigmatism correction, and more.
      • The current role of the MF knobs is displayed in the UI's “Binding” window.
      • MF knobs and programmable buttons can be assigned to various microscope functions, offering versatility and ease of operation.

    Control-panels (left and right)

    Figure 0364d. FEI Microscope Control Panels. Adapted from [1]

 
The typical apertures (4 apertures) included are as follows:
  • C1 - 2000, 70, 50, 30 μm (most common: 2000)
  • C2 - 150, 100, 70, 50 μm (most common: 100)
  • OA - 100, 40, 20, 10 μm (most common: 100)
  • SA - 800, 200, 40, 10 μm
TF20: Tecnai G2 [1]
 
CompuStage: [1]

    CompuStage

  • The left side displays the specimen. Click Add to save the current X-Y stage position, and click Go to return to a previously saved position.
  • On the right side: Control - Adjusts stage movement speed through the Power Step feature; Reset - Resets the XYZ positions and AB tilt angles of the CompuStage to zero. It is recommended to reset the holder (Reset → Holder) both before starting and after completing your work. As an alternative, only XY or tilt (A/B) can be reset.
  • The wobbler is utilized for adjusting the Z-height when introducing a new sample. Note that the button changes to yellow when the wobbler is activated.
 
Alignments:
  • Center the beam: Adjust the Intensity knob to condense the beam to its smallest size. Use the multifunction (MF) knobs for Beam Shift X and Y to center the beam.
  • Center the C2 aperture:
    • The beam should cover the circle evenly.
    • If not, spread the beam slightly beyond the outer circle on the fluorescent screen. Access the aperture control panel (Workset E > Setup E > Motorized Apertures). Check the "Center" box next to "Condenser 2" or click on “Adjust” at "Condenser 2" depending on the specific microscope model.. [6] Use the MF knobs to adjust the position of the C2 aperture. Proper alignment is achieved when the beam spreads evenly around the center of the screen.[9]
  • Condenser 2Apertures

    Figure 0364c. Workset > Tune.

  • C2 condenser astigmatism correction:
    • Choose "Condenser" (the button will highlight in yellow). [11]
    • Choose "Condenser" (the button will highlight in yellow)

      Use the “Multifunction” knobs (both controls pads) to make the beam round; select “None” when finished.
  • The Alignment tab includes two distinct panels: Procedure Alignment and Direct Alignments.
  • Alignments (Procedure Alignment). Procedure Alignments are conducted only when the microscope is significantly misaligned.
  • Direct Alignments. Direct Alignments are typically carried out at the start of each microscope session. The following alignments are typically performed during a TEM session:
    • After each alignment, click “Done” in the Direct Alignments window to reserve the current Gun tilt conditions.
    • Gun Tilt (Optional). Select Gun Tilt in the Direct Alignments menu. Use MF to adjust the beam intensity to its maximum. Then, click the Done button.
      • Press L3 (left panel) or R3 (right panel) to select illumination Spot Size 3.
      • If the beam is off-center, go to Gun Shift in the next step, and use the MF-X and MF-Y knobs to adjust the beam to the center before resuming Gun Tilt alignment.
      • Adjust MF-X and MF-Y to maximize beam Current.
    • Gun Shift.
      • Configure the programmable buttons L3 and R3 for “+ spot size” and “- spot size,” respectively. Adjust the spot size to 3 using these buttons. Condense the beam and select Gun Shift. Center the beam on the screen using the MF knobs. Switch to spot size 9 using the R3 button, condense the beam, and select Align Beam Shift. Again, use the MF knobs to center the beam. Repeat the process until the beam is centered for both spot sizes 3 and 9.
      • Repeat the steps above until both spot sizes beams are centered.
      • Center the other spot sizes in between (4 – 8) using Gun Shift.
    • Beam Tilt (X/Y) (or called condenser deflector balancing).
      • Adjust the microscope to a high magnification (e.g. 500,000x) and locate an empty specimen area.
      • Select "Beam tilt pp X." The beam will start to wobble. Adjust the MF knobs to eliminate the wobbling, ensuring the beam remains stationary at the center of the screen, namely, make the beams overlap. Once aligned, click the "Done" button. Next, choose "Beam tilt pp Y" in the Direct Alignments menu and repeat the same procedure to align the Y pivot points.
    • Beam Shift. Select Beam Shift, and use the MF knobs to align the beam to the center of the screen. Ensure the procedure is completed by clicking the Done button.
    • Rotation Center.
      • Set the microscope at a high magnification (e.g. 500,000x).
      • Re-center the beam on the screen using the trackball.
      • Align the sample using a visible feature as a reference.
      • Expand the beam to illuminate the entire screen area.
      • In the Direct Alignments menu, choose the Rotation Center Alignment option. The feature's image will begin shifting; adjust the movement using the MF knobs to minimize the image movement. Once alignment is complete, click the Done button.
      • Employ high magnification for enhanced precision.
    • Coma-free Pivot Point X and Y.
      • Adjust the microscope to a high magnification, such as 500,000x, and identify a reference object.
      • Converge the beam and adjust the MF-X and MF-Y knobs to make the beams overlap.
    • Coma-free Alignment X and Y. [9]

      Coma-free Alignment X and Y

      • Lift the fluorescent screen and activate the camera view on an amorphous specimen area at approximately 500,000x magnification.
      • Activate the live FFT and adjust the microscope focus until the ring patterns become visible (see page1965).
      • Turn the MF-X and MF-Y knobs to maintain the size and shape of the ring patterns.
    • Astigmatism. [4]
      • Check and correct Monochromator Tune if needed.
      • Select the Stigmators option in the Alignments user interface.
      • Choose the Condenser (LM TEM) option, then balance the Stigmators by shifting the beam to the center from Step 1 to 3 (by clicking the arrows) so that we can ensure the beam is centered to prevent it from shifting while adjusting the Stigmators. Then, adjust the stigmator in Step 4. Here, we need to use the multifunction (MF) knobs to perform the correction until it forms a circle.
      • AstigmatismAstigmatismAstigmatismAstigmatismAstigmatism

      • Do the same thing for Condenser (LM Probe).
      • Do the same thing for Condenser (µP TEM) - microprobe TEM.
      • Do the same thing for Condenser (µP Probe) - microprobe probe mode.
      • Do the same thing for Condenser (nP TEM) - nanoprobe TEM.
      • Do the same thing for Condenser (nP Probe) - nanoprobe Probe.
      • Do the same thing for Condenser (Lorentz TEM).
      • Do the same thing for Condenser (Lorentz Probe).
      • Do the same thing for Objective. Use FFT to check the stigmation. Ensure the granularity in a high-magnification image appears isotropic—using the CCD and live FFT for the best results.
      • CompuStage

      • Do the same thing for Objective 3fdd.
      • Do the same thing for Diffraction. Adjust to make the central spot in the defocused diffraction pattern circular.
      • Save the new alignment file.
    These parameters should be checked regularly to ensure optimal TEM performance. Precise alignment is critical for achieving the best results. Users can select an alignment from the list, adjust it using the MF knobs, and confirm by clicking "Done." Each alignment can be performed individually and repeated as needed. Changes to the alignments will only be saved if the user selects "save changes" when exiting the interface.
  • (Optional) Objective aperture centering:
    • Ensure the beam is centered on the viewing screen and expanded clockwise from the crossover point to form a circle in a proper size on the viewing screen.
    • Press the "Diffraction" button to switch to diffraction mode; the diffraction pattern will appear on the viewing screen. Adjust the camera length as needed using the "Magnification" knob.
    • Position the focusing screen in the beam path, then observe the diffraction pattern through the binoculars. If the diffraction pattern is not centered on the focusing screen, adjust it using the “Multifunction” knobs.
    • Select the desired objective aperture; aperture 4 is recommended for bright-field imaging, while aperture 2 is optimal for high-resolution imaging.
    • Align the objective aperture with the direct beam using the aperture shifting knobs.
    • Choose the "Diffraction" button to switch back to TEM mode.
    • Re-center the beam using the trackball.
  • The alignment files are located in the System drive (for instance, a folder Titan below for a Titan microscope), then locate a folder Alg, where the alignment files with file extension of ".alg" are: [3]

    CompuStage CompuStage

  • The alignment panel is positioned as indicated below, with a list of alignments available in the File tab with the files in the Alg folder:

    CompuStage

    To manage (add, delete, and modify) alignment files, you must log in using either a supervisor or service account. However, normal users can have access to Direct Alignments shown above. And, you can backup the alignment files, and rename the backup files with alignment dates. If alignment files are copied into the Alg folder, they will not appear on the user panel right away. To make them visible on the panel, the server must be restarted.

  • The alignment files can be opened using Notepad:

    CompuStage

 
Monochromator Tune:
  • Locate "Monochromator Tune".
  • Select "Focus" and hold "Shift" to verify the mono-centering.

    Monochromator Tune

    Monochromator Tune

    Monochromator Tune

    Monochromator Tune

 
Taking images with TIA or DigitalMicrograph software:
  • Start the TIA software by clicking on its icon.
  • Navigate stage and find your area of interest.
  • In Microscope Control, select the Camera tab and navigate to the “CCD/TV Camera” control panel.
  • Under “Controller” or "Camera", make sure your desired camera is selected from the pull-down menu.
  • Set “Integration time (s)” = 0.01; then select “Search” (this button and the “Insert” button will turn yellow). Or, pess Insert (a camera, e.g. "BM-Ceta" below) until the button turns yellow as shown in Figure 0364i. In this process, you can listen for the camera to insert.
  • Press R1 to raise the screen.
  • A live image will appear in DigitalMicrograph or Flucam Viewer interface in TIA display once the camera is inserted.

    Ensure that "Tia Ccd" appears in the CCD/TV Camera panel within the Controller window, and "CCD" is displayed in the Camera window
    Cameras
    (a)
    (b)
    (c)
    Figure 0364i. (a) Selecting a Camera and then pressing Insert button, (b) Setting Integration time and then pressing Search, and (c) Yellow colors after pressing Search and Insert.. [1, 11]
  • Adjust the beam focus using the Focus knob and correct the Objective stigmation by selecting “Objective” in the Stigmator window and using the multifunction knobs.
    • In Microscope Control, select the Camera tab; navigate down to the “Stigmator” control panel and select “Objective” (the button will turn yellow).
    • Use the “Multifunction” knobs to adjust the live FFT so the rings are round (see page1965).
    • Click “None” to save.
  • correct the Objective stigmation

  • Ensure that a Camera, e.g. "Tia Ccd" below , appears in the CCD/TV Camera panel within the Controller window, and "CCD" is displayed in the Camera window.

    Ensure that "Tia Ccd" appears in the CCD/TV Camera panel within the Controller window, and "CCD" is displayed in the Camera window

    Figure 0364e. Taking images with TIA software. [1]
  • Select the desired exposure time in the Integration time section.
  • Use the Search, Preview, and Acquire buttons to capture images, with each option allowing different formats and binning settings.
  • MF Knobs can be configured to Image Shift within the Image Settings: [9]
  • MF Knobs can be configured to Image Shift within the Image Settings

  • Select Binning (1 reading per pixel) and Integration Time (1–2 seconds, shorter if the image drifts). Save the images in the Transfer folder as .emi format, or export the image in other formats using right-click. The .emi format contains the information of the experimental condition (magnification, camera length, voltage, etc.)
  • Press “R1” to reinsert the screen and select “Insert” in the CCD/TV Camera window to retract the camera.
  • When performing high-resolution imaging and increasing the magnification, it is advisable to verify the alignment by retracting the objective aperture, refocusing the image, balancing the condenser deflectors, carrying out rotation centering, reinserting and centering the objective aperture in diffraction mode, and correcting objective astigmatism using the live FFT (see Alignments).
  • After using the digital camera, lower the viewing screen by pressing the R1 button on the right-hand control pad.
  • In the Microscope Control interface, select the "Camera" tab, navigate to the "CCD/TV Camera" control panel and click on the "Insert" button to retract the camera (the button will change to gray).

    "Insert" button to retract the camera

 
Taking images with EMMENU4:
  • Launch EMMENU4. [5]
  • EMMENU4

  • Create a new View Port (up to 16 available).
  • Configure the Camera Format (options include 4k x 4k, 2k x 2k, 1k x 1k, with binning levels of 1, 2, 4, or 8).
  • Perform a Flatfield Calibration for each selected format.
  • Select or create an Image Folder and set the Image Name (shortcut: Ctrl+I).
  • Click the Camera Button to capture a single image using the CCD camera, or use the TV Rate Camera Button for continuous image acquisition.
  • For capturing a series of images, adjust the series parameters such as the number of images, delay time, tilt, focus, and gonio alpha.
  • Utilize the comprehensive set of tools available for measurement and image alignment.
  • Save acquired images in either 8-bit or 16-bit format.
 
Loading a specimen into a TEM holder:
  • Close the column valve before sample holder insertion/removal.
  • Reset the stage (holder) before sample holder insertion/removal.
  • TF20 holder: [1]

    TF20 holders

    • The TF20 holders are more sensitive and prone to damage.
    • Inspect the O-ring for any signs of cracks or the presence of dust.
    • Identify the small opening at the bottom of the clamp.
    • Place the tool (pin for lifting sample clamp [9]) into the hole and gently raise the clamp to a vertical position, ensuring it is securely held in place. Avoid using excessive force.

      pin for lifting sample clamp

    • With the clamp in the vertical position, insert the specimen into the carrier.
    • Ensure it is properly centered and gradually lower the clamp into the closed position. Ensure the clamp is fully closed before removing the tool from the hole. The specimen should be positioned with the sample face downward, as the holder rotates approximately 140 degrees during insertion into the column.
 
Load a TEM sample holder into a microscope:
  • Close the column valve before sample holder insertion/removal.
  • Reset the stage (holder) before sample holder insertion/removal. Stage is at the home position: “Search” workset “Stage” pull out menu.
  • Place the specimen into the holder following a proper instruction. Verify that the O-ring and conical surface of the holder are clean and in good condition.
  • Select the tab and then find the “Vacuum” control panel. Ensure the Column Valves are closed (the "Col. Valves Closed" button should appear yellow ).
  • Select the tab and then find the “Vacuum” control panel

  • (For some microscope models) Activate the turbo-molecular pump (TMP) by clicking the Turbo On button. The button's color will transition from gray to orange and finally to yellow. When it turns yellow, the TMP is ready for operation.
  • Identify the small pin positioned near the end of the holder closest to the tip: [1]
  • Identify the small pin positioned near the end of the holder closest to the tip

  • Gently place the holder into the CompuStage (that is, insert the holder straight in to the stop of the stage load lock area), aligning the small pin (the holder guide pin) to the 5 o’clock position (the Close line mark on the goniometer cover) and the large pin near the handle to the 11 o’clock position. [6, 10] Most people encounter a vacuum problem when inserting the TEM sample holder because:
    • The pin is not properly aligned with 'Close.' The red arrow below shows the correct alignment of the pin to the holder stage when inserting the holder as shown in Figure 0364f. (a).
    • The holder has not been pushed into the column with proper pressure after it was inserted.
    If the two key items above are done correctly, the holder insertion status will be as shown in Figure 0364f (b). The turbo pump is turned on, and holder selection button is enabled. If using the double tilt holder, plug in the cable connection on the load lock.
    keep finger pressure on the holder (no rotating!!!) to activate pumping process keep finger pressure on the holder (no rotating!!!) to activate pumping process keep finger pressure on the holder (no rotating!!!) to activate pumping process keep finger pressure on the holder (no rotating!!!) to activate pumping process
    (a)
    (b)
    (c)
    (d)
    Figure 0364f. Holder insertion and pumping procedure: (a) Positions of pin on holder and 'Close' on column. The light is off. (b) Properly inserted holder. The airlock opens with a whooshing sound, and the red light turns on at that moment. (c) The airlockoff. and (d) The final position of the holder. The holder has been turned counterclockwise and pushed in.
    The pre-pumping cycle will begin, and the Red indicator light will illuminate, namely, LED will light up in Red. That is, the red LED on the load lock will illuminate. If the turbo pump is operating at full speed, you should hear a valve open, signaling that the load lock is being pumped down. If the turbo pump is not yet at full speed, the valve will open once it reaches full speed. At this moment, keep finger pressure on the holder (no rotating!!!) to activate pumping process (see the figure above).
  • Wait for the load lock to pump down, until Airlock cycle times out so that the red light turns off (approximately 2-5 minutes; the remaining time can be viewed in the Vacuum Overview - Popup panel). [6] Once the evacuation is complete, the valve will close, and the red LED will turn off as shown in Figure 0364f (c) and the Turbo On is in yellow as shown in Figure 0364f. If the evacuation does not appear to have completed as expected, stop immediately and contact the tool manager for guidance.
  • Airlock cycle times out

    Figure 0364g. Airlock and Turbo pump.
  • Slowly and smoothly rotate the specimen holder counterclockwise until it reaches its limit, namely, its stop position. Next, gently position the holder into the microscope, allowing the column vacuum to securely draw it in.
  • Wait for 1 minute, and then disable the TMP by selecting the Turbo on button, which will change from yellow to gray.
  • Verify the vacuum level in the Column (Workset → Setup → Vacuum). It should be 20 log (10-7 Torr) or lower (see Vacuum).
 
Remove (unload) a TEM sample holder from a microscope:
  • Ensure the instrument remains in SA mode, as confirmed in the bottom information panel.
  • Retract the objective.
  • Close column valves by clicking Col. Valves Closed button (color changes from gray to yellow). The status will now read “COL. VALVES”.
  • Close column valves by clicking Col. Valves Closed button (color changes from gray to yellow!)

  • In the Microscope Control interface, navigate to the "Stage2" control panel and click the flap-out arrow to expand it. Within the expanded panel, locate the "Reset" section and select "Holder" to reset the holder. Ensure the objective aperture is retracted before performing this action; do not initiate the holder reset otherwise.
  • Verify the X, Y, and Z positions and α and β tilt values are all ~0 before proceeding holder removal.
  • Close column valves by clicking Col. Valves Closed button (color changes from gray to yellow!)

  • If using the double-tilt holder, disconnect the β tilt control.
  • While pressing the purple stage cover down with your left hand, carefully pull the holder out of the column until it reaches its limit with your right hand. [10]

    While pressing the purple stage cover down with your left hand, carefully pull the holder out of the column until it reaches its limit with your right hand

  • Turn clockwise until it stops.
  • Wait for 30 seconds.
  • Slowly pull the holder straight out while you are still pushing the purple stage cover down so that you gently remove it from the CompuStage using both hands.
  • In Microscope Control, select the tab and navigate to the “Vacuum” control panel; if the Turbo pump isn’t running, then turn it on (button will be yellow).
 
Shut Down a TEM System:
  • Set the magnification to 5600x.
  • Spread the electron beam.
  • Reset the holder by navigating to Workset → Stage → Stage2 → Control → Reset → Holder. [6]

    Stage → Stage2 → Control → Reset → Holder

  • Close all imaging software, e.g. TIA.
  • Close the column valves by clicking the Col. Valves Closed button in the Setup → Vacuum menu. The button will turn yellow, and the Setup tab will display the status as Column Valves Closed.
  • Take the specimen holder out of the microscope column.
  • If you are the final user of the microscope for the day, ensure you initiate a cryo cycle before leaving. Otherwise, continue to the Cryo Cycle Procedure:
    • Remove the liquid nitrogen dewar from the cold trap.
    • Open the Vacuum tab in the Workset, expand the flap-out menu, and click the Cryo Cycle On button. The TMP will begin operating, and the Cryo Cycle On button will turn yellow.
    In Microscope Control, navigate to the "Startup" tab. In the "Vacuum" control panel, click the flap-out arrow to expand the options, and then access the "Cryo" panel. Select "Cryo Cycle" to begin the cryo cycle. Note that once initiated, the cryo cycle cannot be reversed, and the microscope will remain unavailable for approximately six hours. [11]
  • Select "Cryo Cycle" to begin the cryo cycle. Note that once initiate

  • Close UI window.
  • Log off from TF20 computer.
 
TEM sample holders:
  • Single-tilt: used for morphology analysis and EDS data collection (angle range: -30° to +30°).
  • Double-tilt (low background): suitable for studying morphology, conducting detailed analyses of crystalline materials, and performing EDS acquisition (angle range: -30° to +30°).
  • Tomography (low background): Comparable to a single-tilt holder but designed with a lower profile, enabling a higher alpha tilt angle range (-70° to +70°, depending on the specimen's Eucentric Z setting).
 
Selective Area Diffraction:
  • Obtain a TEM Bright Field image.
  • Choose the necessary field of view.
  • Insert a selected-area aperture with a proper Spot size. [10] Note: Once the Spot size has been changed, the C2 aperture needs to be realigned. Warning: You should not press the diffraction button directly when using a Gatan camera, as the central beam can damage the scintillator of the camera.
  • Insert a selected-area aperture with a proper Spot size

  • Remove the objective aperture.
  • Press the Diffraction button (LED on).
  • Choose the appropriate camera length (magnification) to see the diffraction pattern clearly.
  • Focus the diffraction pattern as sharply as possible to achieve a parallel beam.
  • Adjust the illumination intensity to an appropriate level by turning clockwise to decrease it.
  • Refocus the diffraction pattern if needed.
  • Place the beam stop to block the central diffraction spot to avoid damage to the camera.
  • Place the beam stop to block the central diffraction spot to avoid damage to the camera

  • Remove the objective aperture.
  • Record the diffraction pattern with an appropriate exposure time, balancing the intensity and the risk of damage to the camera.
  • Establish the relationship between the diffraction pattern and the TEM image through orientation calibration which had been done the microscope.
 
Axial dark-field imaging:
  • Center the electron beam.
  • Press Dark-field while in TEM mode.
  • Set the dark-field tilts to 0.00 by pressing Reset.

    Set the dark-field tilts to 0.00 by pressing Reset

  • Acquire the diffraction pattern from the selected region.
  • Align the diffraction pattern on the screen using the diffraction shift adjustment.
  • Determine the Bragg reflection or segment of the polycrystalline ring to use for acquiring a dark-field image.
  • Activate the Dark-Field mode by pressing the Dark field button (the LED will illuminate).
  • Adjust the Multifunction X and Y knobs to align the Bragg reflection opposite the selected one with the point where the central beam was initially positioned.
  • Press Dark-field to switch back to Bright Field Diffraction mode.
  • Position an objective aperture precisely centered on the central spot.
  • Press Dark-field to return to Dark Field Diffraction mode.
  • Precisely align the selected diffraction spot within the objective aperture using the Multifunction X and Y knobs. Ensure the objective aperture is sufficiently small to separate the chosen spot from adjacent diffracted beams.
  • Press the Dark-field button (with the LED off) to display a dark-field image highlighting the crystal planes responsible for the chosen diffraction spot.
  • Perform adjustments to magnification, intensity, beam shift, and focus similar to those used for a Bright Field image.
 
Off-axis dark-field imaging:
  • Press diffraction button.
  • Center the objective aperture around the diffracted beam needed for dark-field imaging. Ensure the objective aperture is sufficiently small to separate the chosen spot from adjacent diffracted beams.
  • Switch back to imaging.
 
Finish a microscope session:
  • Stop Camera View (search) and Retract CCD (BM-ceta) camera by clicking the Insert camera again when the camera is in.
  • Select Col Valves Closed under Setup > Vacuum to close the column valves.
  • Press the R1 button to lower the fluorescence screen, then place the rubber cover on.
  • Navigate to the Reset panel under Search → Stage → Control → Reset page, and select Holder to center the stage.
  • While holding down the purple stage cover, gently pull the holder straight out until it stops.
  • Turn the holder clockwise until it stops.
  • Wait for 30 seconds.
  • While keeping the purple stage cover pressed down, carefully pull the holder straight out.
  • Remove and save the specimen.
  • Insert the empty single tilt holder in to goniometer. The single-tilt holder should remain inserted in the column when the TEM is not in active use to maintain the cleanliness of the load lock area and column.
  • Log off your microscope session via FOM.
 
 
 
 
 

 

 

 

 

 

[1] https://eicn.cnsi.ucla.edu/wp-content/uploads/2018/01/TF20Manual-LastVersion.pdf.
[2] https://amcc.mst.edu/.
[3] N. Rudawski, FEI S/TEM alignment file basics, 2024.
[4] N. Rudawski, FEI Themis Z S/TEM: Stigmators, 2024.
[5] https://www.tvips.com/imaging-software/em-menu/.
[6] http://www.nano.pitt.edu/sites/default/files/Equipment-SOP/FEI%20Titan%20Themis%20G2%20200%20Operations%20Manual%20TEM%20Rev-5.pdf.
[7] N. Rudawski, FEI S/TEM FEG Registers: Uses and Cautions, 2020.
[8] https://zmb.dozuki.com/Guide/TEM+-+FEI+Tecnai+Spirit/173.
[9] https://static1.squarespace.com/static/57b26cc76b8f5b7524bf9ed2/t/5955124b2994caad3a3e37cf/1498747472510/TEM_Talos_SOP_CU_June2017.pdf.
[10] N. Rudawski, FEI Tecnai F20 S/TEM: selected area diffraction, 2019.
[11] https://nrf.aux.eng.ufl.edu/_files/documents/933.pdf.